Medical machine for purified water
Water supply to the cleaning room:
It is mainly intended for cleaning using clean water for automatic washing machines, final cleaning, steam generators, etc. The electrical resistance of the water for the washing machine and final cleaning should be less than 15 µcm / cm, and for the steam generator — less than 5 µcm / cm.
Water for cleaning soft endoscopes (including gastroscopy and colonoscopy):
The main indicators are that the electrical resistance of water should be < 15 microns/cm, the colony content < 10 CFU/100 ml, the pore size of the filter membrane during the production of purified water should be < 0.2 microns, and it should be replaced regularly.
Use an osmosis system depending on the water quality in your area. If the water quality is more severe, use a two-stage osmosis system, if the water is better, use a single-stage osmosis system.